U8557560-230 - Sdílet Measurement Amplifier U (230 V, 50/60 Hz)
U8557560-230 - Sdílet Measurement Amplifier U (230 V, 50/60 Hz)
Order Number: 5401.1020742
Measurement amplifier U amplifies low-amplitude measurement signals from low-resistance signal sources for measurement with any chosen voltmeter or oscilloscope.
More than a purchase. You get service and expert advice. Ask which products and combinations are recommended for you.
Measurement amplifier U amplifies low-amplitude measurement signals from low-resistance signal sources for measurement with any chosen voltmeter or oscilloscope. By using an external shunt resistor it is also possible to measure small currents. Offset voltages can be compensated using coarse and fine offset adjustment knobs. Amplification (gain) can be selected in ranges from 0 to 5 powers of ten. High-frequency noise or other interference signals are filtered out by means of a low-pass filter with step-wise selectable time constants between 0 and 3 seconds. The output voltage has the same sign as the input voltage.
Technical Data:
Input resistance: 10 kohm
Output resistance: 300 ohm
Offset voltage drift: < 2 μV/K (after 15 mins.
operation approx.)
Gain factors: 100; 101; 102; 103; 104; 105
Tolerance for gain factors: < 2.5 %
Input voltage: max. ±12 V
(overload protected for
brief transients up to 100 V)
Output voltage: 0 ... ±12 V
(short-circuit protected)
Power supply (via
plug-in supply provided): 12 V AC
Ambient temperature: 5°C ... 23°C ... 40°C
Storage temperature: -20 ... 70°C
Relative humidity: <85%
no condensation
Operational alignment: Horizontal
Contamination level: 2
Protection class: IP20
Dimensions: 170x105x50 mm3 approx.
Weight: 335 g approx.
Measurement and frequency ranges (relative to 1 V output voltage)
Gain
Input voltage
Current through RShunt = 100 ohm
Frequency
100
1 V
10 mA
0 ... 25 kHz
101
100 mV
1 mA
0 ... 25 kHz
102
10 mV
100 µA
0 ... 25 kHz
103
1 mV
10 µA
0 ... 20 kHz
104
100 µV
1 µA
0 ... 7 kHz
105
10 µV
100 nA
0 ... 7 kHz
Cut-off frequency for switchable low-pass filter
Time constant
Cut-off frequency
0.0 s
See Table 1
0.1 s
1.6 Hz
0.3 s
0.5 Hz
1.0 s
0.16 Hz
3.0 s
0.05 Hz
Devices - filter
Devices for measuring power and energy
Quantitative unit
ks