More than a purchase. You get service and expert advice. Ask which products and combinations are recommended for you.
XS Instruments PC 50 VioLab DHS Bench multimeter
pH – mV/ORP – COND – Temperature
Large backlight display pH/mV/Conductivity and Temperature parameters simultaneously
pH
Selectable resolution: 0.1/0.01 pH
Automatic calibration up to 3 points with 2 selectable pH-buffer sets (USA and NIST)
Manual calibration up to 2 points with user-defined pH-Buffer
Stability parameter (Low-Med-High) for high accuracy measurement
Conductivity
Automatic calibration from 1 to 4 points between 4 conductivity standard values: 84 µS – 1413 µS – 12.88mS – 111.9 mS
Manual calibration at 1 point (User defined)
Automatic and manual compensation of temperature with reference value 15…30 °C
PC 50 VioLab DHS complete kit include pH electrode 201T DHS and conductivity cell 2301T with integrated ATC, both with 1m cable and BNC plug, electrode holder and colored pH buffer solution pH 7 and pH 4, standard calibration solutions 1413 µS and 12,88 mS and 5V/220V AC power supply.
pH
Measuring range
0.00…14.00pH
Resolution
0.1/0.01pH
Calibration points
1, 2 or 3
Buffer recognized
USA: 1.68 – 4.01 – 7.00 – 10.01 pH; NIST: 1.68 – 4.01 – 6.86 – 9.18 pH; 2 points defined by user
Indication of the Slope and Offset
Yes
Criteria of measurement stability
Yes
Indication of calibration point
Yes
Potencial measuring range
± 1.000 mV
Potencial resolution
1 mV
Conductivity
Measuring range
0.0…200 – 2000 µS/20.00–200.0 mS
Resolution
0.1/1 µS to 0.01/0.1 mS
Relative accuracy
±2% full scale
Calibration points
1..4
Calibration standards recognized
84µS, 1413µS, 12.88mS, 111.9mS, 1 point user defined
Cell constant selectable
0.1, 1, 10 cm-1
Temperature coefficient
0.00…10.00 %/°C
Refenece temperature
15…30 °C
Indication of calibration standards
Yes
Measuring range
0…100 g/l
TDS factor/Accuracy
0,40..1,00/±1% full scale
Temperature
Measuring range
0…100.0°C
Resolution/Accuracy
0.1°C / ±0.5°C
Temperature compensation pH
Automatic/manual 0…100°C
Temperature compensation COND
Automatic/manual 0…80°C
Others
Inputs
2 BNC, 2 RCA-CINCH (ATC)
Power supply
5V DC/220 V
Dimensions/weight instrument
160×185×70(H)mm/550g
Use
MultiparameterPhys. property
StolníInstrument type
Quantitative unit
ks
PC 50 VioLab kit with pH electrode CHS ChemFlex S7, cond.cell CHS CondiLab 22P, standards and access.