Temperature range: 0 °C to 70 °C
Temperature range with light: 10 °C to 60 °C
Humidity range: 10 % to 80 % RH
3 positionable illumination cassettes with ICH-compliant UV/Vis light source
Independent light-dose control of UV-A and visible light with spherical sensors
APT.line™ preheating chamber technology
Humidity regulation with capacitive humidity sensor and vapor humidification
Inner chamber made of stainless steel
Intuitive touchscreen controller with time-segment and real-time programming with internal data recording
Tight-sealing inner door made of safety glass (ESG) with special TIMELESS coating
3 stainless steel racks
Access port with silicone plug, 30 mm, left
4 stable castors, two with brakes
Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
Computer interface: Ethernet
Door heating
Constant climate chambers with ICH-compliant light source and light dose control
One special advantage of BINDER's KBF LQC series of Constant climate chambers is the light photometry, an enhancement for photostability testing developed and patented by the company. This involves two spherical ball-type sensors capturing the light intensity at the sample location. Once the set light value has been achieved, the light source switches off automatically. The ball-type sensors can be flexibly placed on the sample material. The Constant climate chambers in the KBF LQC series use light dose control to enable you to work with precision and in compliance with standards.
In addition to accurate light photometry, BINDER Constant climate chambers also provide precise climate conditions thanks to the patented APT.line™ preheating chamber technology, which ensures that air is circulated extremely evenly, even when the chambers are fully loaded. Homogeneous climate conditions are guaranteed across the entire usable space.
Constant climate chambers with homogeneous light distribution in the usable space.
BINDER Constant climate chambers with ICH-compliant light source and light dose control have a unique light concept that guarantees homogeneous light distribution. The illumination cassettes can be positioned variably and even removed. The chambers' humidity system is based on a capacitive humidity sensor and vapor pressure humidification, which offers quick response times. These Constant climate chambers also feature an automated water and waste water management system.
BINDER Constant climate chambers with ICH-compliant light source and light dose control are solutions boasting state-of-the-art technology, with maximum precision and functionality guaranteed. These incredibly versatile units expertly combine performance and functionality with a durable quality and ease of use.